High-Precision Digital Density Meter DA-860
● Kyoto Electronics Manufacturing Co., Ltd. (KEM) is a manufacturer of density standard liquids in Japan compliant with ISO 17034, traceable to national standards.
● KEM's oscillating-type density meters are calibrated using JCSS laboratory density standards compliant with the International MRA and ISO 17025, traceable to international standards worldwide.
● High accuracy: 3×10⁻⁶ g/cm³, high repeatability: 1×10⁻⁶ g/cm³, high reproducibility: 2×10⁻⁶ g/cm³ (under ideal conditions according to ISO 5725).
High-Precision Digital Density Meter DA-860 Key Features:
1. High reliability: Internationally certified manufacturer of density standard liquids ensures instrument accuracy.
2. Equipped with camera function: Enhances visibility of the measurement unit, displays real-time images, and allows storage.
3. Optional operation unit: Choose between a tablet or computer based on your analysis style.
4. Reduced measurement time: Achieves four-digit accuracy in 10 seconds through advanced algorithms.
5. Multi-channel measurement: The operation unit can control multiple measurement units for convenient data management.
6. Comprehensive functions: Automatic bubble detection, full-range viscosity correction, built-in atmospheric pressure correction.
7. Fully automatic cleaning: One-touch operation for automatic sampling, measurement, drainage, cleaning, and drying.
8. Accuracy: 3×10⁻⁶ g/cm³, repeatability: 1×10⁻⁶ g/cm³, reproducibility: 2×10⁻⁶ g/cm³.
High-Precision Digital Density Meter DA-860 Technical Specifications:
Measurement range: 0–3 g/cm³.
Measurement temperature: 0–100°C (32–212°F), with built-in Peltier temperature control.
Accuracy: Density: 3×10⁻⁶ g/cm³.
Repeatability: Density: 1×10⁻⁶ g/cm³.
Reproducibility: Density: 2×10⁻⁶ g/cm³, temperature: 0.001°C.
Measurement method: U-shaped oscillating tube method.
Sample requirement: Manual injection approx. 1 mL, automatic pump injection approx. 2 mL.
Measurement time: Approx. 10 seconds under stable temperature conditions.
Display: Via tablet or computer.
Viscosity correction: Automatic viscosity correction function.
Sampling method: Manual injection or automatic sampling via built-in injection pump.
Measurement cell check function: Automatically checks the dryness of the measurement cell; the pump stops automatically when the cell is dry.
Stability judgment: Five-stage adjustment based on measurement accuracy and time.
Concentration conversion: Automatically converts concentration values by inputting density and concentration tables or conversion formulas.
Temperature compensation: Automatically performs temperature compensation by inputting density and temperature tables or conversion formulas.
Statistical calculation: Automatically or manually calculates average, standard deviation, and relative standard deviation.
Power supply: AC100–240V, 50/60Hz, 40W.
Dimensions: 295(W)×330(D)×255(H) mm.
Weight: Approx. 19 kg.
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 京都电子KEM |
| Spec: | DA-860 |
| Stock: | |
| Origin: |